/*! \page page_calibration Device Calibration
|
|
|
|
\tableofcontents
|
|
|
|
\section calibration_self Self-Calibration
|
|
|
|
UHD software comes with several self-calibration utilities for
|
|
minimizing IQ imbalance and DC offset. These utilities perform
|
|
calibration sweeps using transmit leakage into the receive path (special
|
|
equipment is not required). The results from a calibration are written
|
|
to a CSV file in the user's home directory. UHD software will
|
|
automatically apply corrections at runtime when the user re-tunes the
|
|
daughterboard LO. Calibration results are specific to an individual RF
|
|
board.
|
|
|
|
<b>Note:</b> When a calibration table is present, and the user wishes to
|
|
override the calibration settings through the API: the user should
|
|
re-apply the desired setting every time the LO is re-tuned.
|
|
|
|
UHD software comes with the following calibration utilities:
|
|
|
|
- **uhd_cal_rx_iq_balance:** - minimizes RX IQ imbalance vs. LO
|
|
frequency
|
|
- **uhd_cal_tx_dc_offset:** - minimizes TX DC offset vs. LO
|
|
frequency
|
|
- **uhd_cal_tx_iq_balance:** - minimizes TX IQ imbalance vs. LO
|
|
frequency
|
|
|
|
The following RF frontends are supported by the self-calibration
|
|
utilities:
|
|
|
|
- RFX Series transceiver boards
|
|
- WBX Series transceiver boards
|
|
- SBX Series transceiver boards
|
|
- CBX Series transceiver boards
|
|
- UBX Series transceiver boards
|
|
|
|
\subsection calibration_self_utils Calibration Utilities
|
|
|
|
UHD software installs the calibration utilities into
|
|
`<install-path>/bin`. **Disconnect** any external hardware from the
|
|
RF antenna ports, and run the following from the command line. Each
|
|
utility will take several minutes to complete:
|
|
|
|
uhd_cal_rx_iq_balance --verbose --args=<optional device args>
|
|
uhd_cal_tx_iq_balance --verbose --args=<optional device args>
|
|
uhd_cal_tx_dc_offset --verbose --args=<optional device args>
|
|
|
|
See the output given by `--help` for more advanced options, such as
|
|
manually choosing the frequency range and step size for the sweeps.
|
|
|
|
<b>Note:</b> Your daughterboard needs a serial number to run a calibration
|
|
utility. Some older daughterboards may not have a serial number. If this
|
|
is the case, run the following command to burn a serial number into the
|
|
daughterboard's EEPROM:
|
|
|
|
<install dir>/lib/uhd/utils/usrp_burn_db_eeprom --ser=<desired serial> --args=<optional device args>
|
|
|
|
\subsection calibration_data Calibration Data
|
|
|
|
By default, calibration files are stored in the user's home/application
|
|
directory:
|
|
|
|
- **Linux:** `${HOME}/.uhd/cal/`
|
|
- **Windows:** `%APPDATA%\.uhd\cal\`
|
|
|
|
If you would like to specify a custom directory, you can do so with the
|
|
`$UHD_CONFIG_DIR` environment variable:
|
|
|
|
- **Custom (any OS):** `${UHD_CONFIG_DIR}/.uhd/cal/`
|
|
|
|
Calibration files can easily be moved from one machine to another by copying the
|
|
"cal" directory. Re-running a calibration utility will replace the existing
|
|
calibration file. The old calibration file will be renamed so it may be
|
|
recovered by the user.
|
|
|
|
|
|
\subsection ignore_cal_file Ignoring Calibration Files
|
|
|
|
At runtime, the user can choose to ignore a daughterboard's calibration file by
|
|
adding "ignore-cal-file" to the arguments. With the UHD API, it can be done as
|
|
follows:
|
|
|
|
~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~{.cpp}
|
|
uhd::device_addr_t hint;
|
|
hint["type"] = "usrp1";
|
|
hint["ignore-cal-file"] = "";
|
|
uhd::device_addrs_t dev_addrs = uhd::device::find(hint);
|
|
~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
|
|
|
|
Using tx_waveforms as an example, the user can apply this argument as follows:
|
|
|
|
tx_waveforms --args="addr=192.168.10.2,ignore-cal-file" --freq=100e6 --rate=1e6
|
|
|
|
*/
|
|
// vim:ft=doxygen:
|